On the possibility to improve the performance of flow injection analysis by deconvolution of spectrophotometric data

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

The concentration profile of chemical species pumped through a tubular system of 0.5-mm inner diameter is found to be trapezial shaped, as described theoretically by including the spatial resolution of the instrument, friction and diffusion. By applying the theory, which is based on the concept of segment flow, to experiments, it is shown that an injected segment of solute is diluted predominantly at the front and at the tail proportion of the segment in motion. A double-peaked structure in experiments of flow-injection analysis (FIA) was identified and by including all the data of the experimentally measured responses, it was possible to obtain an improvement of the sensitivity by a factor of 22 and obtain almost a factor of three times improvement on the standard deviation.

Original languageEnglish
Pages (from-to)131-138
Number of pages8
JournalTalanta
Volume54
Issue number1
DOIs
Publication statusPublished - Mar 30 2001

Fingerprint

Flow Injection Analysis
Friction
Deconvolution
Experiments

All Science Journal Classification (ASJC) codes

  • Chemistry(all)

Cite this

@article{1f428b58bf1b4e41abe4628749eb9e19,
title = "On the possibility to improve the performance of flow injection analysis by deconvolution of spectrophotometric data",
abstract = "The concentration profile of chemical species pumped through a tubular system of 0.5-mm inner diameter is found to be trapezial shaped, as described theoretically by including the spatial resolution of the instrument, friction and diffusion. By applying the theory, which is based on the concept of segment flow, to experiments, it is shown that an injected segment of solute is diluted predominantly at the front and at the tail proportion of the segment in motion. A double-peaked structure in experiments of flow-injection analysis (FIA) was identified and by including all the data of the experimentally measured responses, it was possible to obtain an improvement of the sensitivity by a factor of 22 and obtain almost a factor of three times improvement on the standard deviation.",
author = "Andersen, {Jens E.T.}",
year = "2001",
month = "3",
day = "30",
doi = "10.1016/S0039-9140(00)00642-1",
language = "English",
volume = "54",
pages = "131--138",
journal = "Talanta",
issn = "0039-9140",
publisher = "Elsevier Science Publ Co Inc",
number = "1",

}

On the possibility to improve the performance of flow injection analysis by deconvolution of spectrophotometric data. / Andersen, Jens E.T.

In: Talanta, Vol. 54, No. 1, 30.03.2001, p. 131-138.

Research output: Contribution to journalArticle

TY - JOUR

T1 - On the possibility to improve the performance of flow injection analysis by deconvolution of spectrophotometric data

AU - Andersen, Jens E.T.

PY - 2001/3/30

Y1 - 2001/3/30

N2 - The concentration profile of chemical species pumped through a tubular system of 0.5-mm inner diameter is found to be trapezial shaped, as described theoretically by including the spatial resolution of the instrument, friction and diffusion. By applying the theory, which is based on the concept of segment flow, to experiments, it is shown that an injected segment of solute is diluted predominantly at the front and at the tail proportion of the segment in motion. A double-peaked structure in experiments of flow-injection analysis (FIA) was identified and by including all the data of the experimentally measured responses, it was possible to obtain an improvement of the sensitivity by a factor of 22 and obtain almost a factor of three times improvement on the standard deviation.

AB - The concentration profile of chemical species pumped through a tubular system of 0.5-mm inner diameter is found to be trapezial shaped, as described theoretically by including the spatial resolution of the instrument, friction and diffusion. By applying the theory, which is based on the concept of segment flow, to experiments, it is shown that an injected segment of solute is diluted predominantly at the front and at the tail proportion of the segment in motion. A double-peaked structure in experiments of flow-injection analysis (FIA) was identified and by including all the data of the experimentally measured responses, it was possible to obtain an improvement of the sensitivity by a factor of 22 and obtain almost a factor of three times improvement on the standard deviation.

UR - http://www.scopus.com/inward/record.url?scp=0035970944&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0035970944&partnerID=8YFLogxK

U2 - 10.1016/S0039-9140(00)00642-1

DO - 10.1016/S0039-9140(00)00642-1

M3 - Article

AN - SCOPUS:0035970944

VL - 54

SP - 131

EP - 138

JO - Talanta

JF - Talanta

SN - 0039-9140

IS - 1

ER -